89
12.3
Operations for Syringes
12.3.1 Removing a Syringe
12.3.2 Changing a Syringe
12.3.3 Syringe Replacement Interval
1.
Press
to stop the infusion.
2.
Disconnect the patient.
3.
Open the syringe barrel clasp.
4.
Press
to silence the audible signal for 2 minutes.
5.
Push the disengagement lever down and remove the syringe from its cradle.
6.
Disconnect the syringe from its extension set.
7.
Disconnect the extension set from the access device in accordance with
healthcare facility protocol.
1.
Remove the syringe,
2.
Prepare a new syringe and follow the steps described in the flowchart.
3.
After the new syringe is installed, acknowledge the "same therapy" screen
(optional screen).
Warning
Electronically prime the syringe pump system after replacing a near-
empty syringe with a replacement syringe.
Information
Properly dispose of used syringes.
Replace the syringe according to your healthcare facility’s protocol or CDC
guidelines.