109
Figure 17.4:
Trumpet curves for 2, 5, 11, 19, 31 minutes observation windows
(1 mL/h over first 2 hours on 96 hours)
17.9.3 Flow rate: 5 mL/h
Figure 17.5:
Start-up and instantaneous flow rate
(5 mL/h over first 2 hours on 96 hours)
Figure 17.6:
Trumpet curves for 2, 5, 11, 19, 31 minutes observation windows
(5 mL/h over first 2 hours on 96 hours)
Legend
Measured
variance from
flow rate
Error
Flow rate
Sampling time
: 10 s
Legend
Instantaneous
flow rate
Set flow rate
Sampling time
: 10 s
Legend
Measured
variance from
flow rate
Error
Flow rate
Sampling time
: 10 s
F
low r
ate
(m
L/h)
Time (minutes)
6,0
3,9
1,0
1,1
-8,6
-6,5
-4,4
-1,9
-1,6
-10
-5
0
5
10
0
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30
F
low r
ate
(m
L/h)
Time (minutes)
0
1
2
3
4
5
6
7
8
9
10
0
10
20
30
40
50
60
70
80
90
100
110
120
F
low r
at
e
(m
L/h
)
Time (minutes)
1,7
1,0
0,7
0,4
0,2
-1,9
-1,0
-0,7
-0,5
-0,3
-10
-5
0
5
10
0
2
4
6
8
10
12
14
16
18
20
22
24
26
28
30