78
9
Data Communication
9.1
Overview
9.2
Communication via Agilia Cables
9.2.1
Data Communication Cables
9.2.2
Using the Communication Port
Cable Communication
Wi-Fi Communication
Connection of 1 pump to a PC for the following
purposes:
Data set upload
Maintenance
Communication between a hospital information
system server and a number of identified pumps
for the following purposes:
Data set upload
Pump history retrieval
Information
Ensure that all hospital information systems have been approved
by
Fresenius Kabi
. For more information, contact your technical
services representative.
Before connecting the pump to a hospital information system, ask
your IT or biomedical department to configure the device.
Information
Only use recommended Agilia cables.
All connections and disconnections must be performed by
qualified and appropriately trained staff.
All IT devices (including computers, hubs and switches) inside the
patient area (< 4.92 ft or < 1.5 m) must comply with IEC/EN
60601-1 (leakage current).
IT devices connected outside the patient area (> 4.92 ft or
> 1.5 m) must be at least IEC/EN 60950 compliant.
1.
Remove the protective cap from the pump’s
RS232 communication port.
2.
Connect the cable to the RS232 communication
port by turning the cable wheel.