45
6.8.3.1
Direct Bolus
6.8.3.2
Programmed Bolus
Information
The bolus volume is added to the Volume Infused (VI).
The
key is not active when the menu screen is displayed.
Note: This feature can be activated or deactivated with a compatible DERS
(custom profiles) or in the pump options (Basic profile).
1.
During the infusion, press
.
2.
Press
bolus
to confirm access to the Direct Bolus function.
3.
To administer a direct bolus, press and hold
.
4.
Monitor the volume infused on the main display until the desired bolus is
reached.
5.
To stop the bolus, release the
key.
The infusion resumes its previous rate after the bolus is delivered.
Note: This feature can be activated or deactivated with a compatible DERS
(custom profiles) or in the pump options (Basic profile).
During the infusion, you can program a bolus in one of the following two ways:
Press
, then
prog
.
Press
, and select
in the menu. Press
enter
to confirm.