58
7.4
Keypad Lock Status
Information
The Dynamic Pressure System (DPS) informs the user of any
sudden rise or drop in pressure before the pressure threshold is
reached.
If variable pressure mode is enabled, a pre alarm is triggered
when the pressure reaches 50 mmHg below maximum pressure
(25 mmHg when maximum pressure is 50 mmHg).
If other pumps are used in parallel, it is recommended that their
pressure thresholds be adjusted to the same level.
Symbol
Procedure
Locking / Unlocking the keypad
You can use this feature to avoid inadvertent key presses.
Note: The following features can be activated or deactivated in the pump options:
Automatic lock: The keypad will lock automatically at infusion start, or after
a time-out.
Unlock code: The user must enter a code to unlock the keypad.
Locking the Keypad
You can lock the keypad as follows:
1.
Press
.
2.
Press the arrow keys to select
.
3.
Press
enter
.