83
11.4.3 Volume/Time & Dose/Time Alarms
11.4.4 Infusion Alarms
Near end of volume
limit !!
Medium (!!)
No
The time remaining is less than the defined
time duration (adjustable between 1 and
30 minutes), and the remaining VTBI until the
volume limit has dropped to less than 10% of
the syringe capacity.
End of volume
limit !
Low (!)
No
The volume limit is reached and the end of
infusion setting is set as "KVO" or
"continuous".
Message
Priority
Stops
Infusion?
Problem / Resolution
End of
volume/time !!!
High (!!!)
Yes
The VTBI is completed.
End of dose/time !!!
Near end of
volume/time !!
Medium (!!)
No
The time remaining is less than the defined
time duration (adjustable between 1 and
30 minutes), and the remaining VTBI has
dropped to less than 10 % of the syringe
capacity.
Near end of
dose/time !!
End of volume/time !
Low (!)
No
The VTBI is completed and the end of infusion
setting is set as "KVO" or "continuous".
End of dose/time !
Message
Priority
Stops
Infusion?
Problem / Resolution
End of infusion !!!
High (!!!)
Yes
The infusion is completed (simple rate).
Near end of
infusion !!
Medium (!!)
No
The time remaining is less than the defined
time duration (adjustable between 1 and
30 minutes), and the remaining volume of fluid
in the syringe has dropped to less than 10 %
of the syringe capacity.
Check settings !!
Medium (!!)
No
The flow rate (or dose rate) has been modified
using the keys, but has not been confirmed.
Check the flow rate (or dose rate) and
press
OK
to confirm.
Waiting settings !!
Medium (!!)
No
A value must be entered.
Enter a value and press
OK
to confirm.
Waiting start !!
Medium (!!)
No
The infusion settings have been entered, but
have not been confirmed with
start
.
Check the infusion settings, and
press
start
to start the infusion.
Message
Priority
Stops
Infusion?
Problem / Resolution