12
1.7
Contraindications
1.8
Use Environment
Do not modify the pump (except in the case of operations recommended
by
Fresenius Kabi
).
Do not use the pump with the following
fluids
:
- Flammable liquids
- Fluids not suitable for infusion
Do not use the pump in the following
environments
:
- Explosive or flammable environments
- High humidity environments (shower, bath, etc.)
- Ultrasonic environments
- Magnetic Resonance Imaging (MRI)
- Hyperbaric chamber
Do not use the pump for the following
purposes
:
- Infusion in association with a dialyser or ECMO
- Enteral nutrition
- Epidural use
Do not allow the pump to come in direct contact with the patient's body.
While the pump is infusing a patient, do not connect a computer installed
with Agilia Partner software to perform technical operations.
Agilia SP MC is intended for use in healthcare facilities, under the supervision of
trained healthcare personnel.
The pump must be used in the following operational conditions to ensure proper
performance:
Operating temperature range:
41 °F (5 °C) to 104 °F (40 °C)
Operating pressure range:
700 hPa (525 mmHg / 10.15 PSI) to 1060 hPa (795 mmHg / 15.37 PSI)
Operating humidity range:
20 % to 90 % with no condensation
Altitude:
Up to 9842.52 ft (3000 m) above sea-level
Warning
The functionality of the pump can be affected by pressure variations,
mechanical shocks, heat ignition sources, and so on.
Information
For more information on using the device in specific conditions, contact
your
Fresenius Kabi
sales representative.