108
17.9.1 Flow rate: 0.1 mL/h
Figure 17.1:
Start-up and instantaneous flow rate
(0.1 mL/h over first 20 hours on 96 hours)
Figure 17.2:
Trumpet curves for 4, 20, 60, 200, 320 minutes observation windows
(0.1 mL/h over first 20 hours on 96 hours)
17.9.2 Flow rate: 1 mL/h
Figure 17.3:
Start-up and instantaneous flow rate (1 mL/h over first 2 hours on
96 hours)