74 / 78
6
Transportation and installation of radiation source module
The radiation source will be assembled into the radiation source module at the
manufacturer’s site and will be supplied as calibrated and tested assembly. Separate from
the
lexsyg
main unit, the radiation source module is being shipped in a certified type A
shipping container (fig. 10, attachment 2).
Fig. 10
Type A shipping container (left) and source container with radiation source
module (right)
The radiation source module can be easily installed into the
lexsyg
per instructions provided
below and depicted by figures 10
– 12.
-
Upon receipt of radioactive source shipping perform activities as required by local
radiation policy involving radiation survey, wipe test on the shipping container,
document check etc.
-
Open shipping drum and remove upper part of the Styrofoam insert.
-
Remove source container from Styrofoam insert.
-
Open source container.
-
Remove radiation source module from source container. Ensure transport safety
plate is properly affixed to the radiation source module.