3.6.4.
Initialization after an Incident
– First aid menu
If an uncommon event occurs during a measurement, it is necessary to reinitialize the
system. The first step is to click the
Reset Lexsyg and initialize new
button (Figure 22
–1)
in the first aid menu. Wait until the
measurement
button changes its caption to
start
measurement
. If no sample disc is remaining inside the measurement chamber, a new
measurement can be started.
If there is still a sample disc in the measurement chamber, this sample disc has to be
transferred back to the sample wheel before proceeding with measurements. Therefor
localize an empty sample slot in the sample wheel and open the
First aid
menu (Figure 22).
If there is still a sample disc is in the grabber and not on the sample arm, or vice versa,
check the radio buttons
sample on arm
or
sample on grabber
– 4)
corresponding to the sample disc location. Then Click the button
Send new sample
allocation
– 2)
.
After that, select an empty slot which has to be filled with the
sample disc from the measurement chamber (Figure 22
– 5) and click
Try to get sample
from arm
or
Try to get sample from grabber
– 3).
Figure 22: first aid menu
The sample disc is now transferred from the corresponding location to the sample wheel. If it
is unknown, if a sample disc is located in the grabber or on the sample arm after the re-
initialization, begin with transferring back from the grabber and after that with transferring
back from the sample arm to an additional free sample wheel slot.