no stray-light transmission outside the optical path
optimized filter-sets for all main quartz and feldspar minerals luminescence emission
and other materials/applications available
filter sets (interference, colour-glass optical filters) optimised for highest transmission
individual wave-bands and best blocking stimulation light (OSL) and/or thermal noise
(TL, black-body radiation)
modified detection windows can be ordered (includes software-adaption)
malfunction / PMT-damage software protected
2.1.11.
Module 6.1.3: Reduced PMT optics unit
These optical modules are designed for use with PMT units only. They are best performed to
collect the luminescence light from the sample, suppress stray-light (OSL, PL) or thermal
noise (TL), filter it and transfer the light efficiently to the photo-cathode of the PMT.
configuration to perform optimized TL/OSL analyses in dating and other applications
optics designed for high-efficiency PMT luminescence measurements
1measurement function of 6.1.2 is included (more information on request)
malfunctions / PMT-damage are software protected
no stray-light transmission outside the optical path
optimized filter-sets for all main quartz and feldspar minerals luminescence emissions
and such for other materials/applications are available
2.1.12.
Module 6.2.1: Imaging optics adaption & filter
changing unit for TL/OSL measurement position
The
lexsyg
can adapt imaging light detectors based on charged coupled devices; in
particular an EMCCD based module (5.2.1) is available for spatially resolved measurements,
incl. single-grain measurements. Such detection units need advanced optical design for best
imaging quality within a chosen wavelength-band to be measured which is set by the optical
filters detection- window. Imaging units also work with PMT detectors and thus assure best
performance if a LEXSYG measurement chamber with automatically detector change is in
use.
necessary for spatially resolved / single-grain TL/OSL measurements