![Freiberg Instruments lexsyg research Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/freiberg-instruments/lexsyg-research/lexsyg-research_manual-and-documentation_2331334052.webp)
Freiberg
Instruments
GmbH
ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
52 / 78
figure 39:
maintenance panel for changing the heater unit
Now the measurement chamber has to be moved up (see point 4.1.2).
The arm can be removed by the three screws in the centre.
figure 40:
removedmaintenance panelwith visibleheater unit
Then the arm can be moved up carefully and a new one can be mounted in reverse direction.
4.2.
Spare parts
Spare parts:
-
Chamber window glass
-
Gasket for all removeable parts connected to the chamber
-
Screws
-
Air tubes
-
Wires and plugs
-
Sample arm with heater
Tools:
-
Cleaning brush
-
Filter sucker
-
Set of Allen wrench
Requests and feedback is welcome, please contact:
Open the panel by unmounting the
screws