Maintenance and service
The radiation source is exclusively used under laboratory conditions. It is not subjected to
any excessive mechanical, thermal or chemical loads other than specified in the appropriate
ISO 2919 testing protocol.
The sealed beta sources of type Sr0.S03-1 are embedded in a stainless steel holder and
protected by a cover foil (25 µm stainless steel), see 8. The special design of the source
holder as well as regularly conducted visual inspections and leakage tests (refer to section 5)
ensure radiation source integrity at any time.
The radiation source module, in particular the shutter mechanism, has undergone extensive
testing and functional performance verification. Anticipating the shutter to be activated less
than 1000 (one thousand) times annually, the manufacturer recommends an annual
inspection and functionality verification of the radiation source module. The inspection is to
be carried out by trained personnel only per instructions provided by the manufacturer.
The radiation source has a recommended working life of 10 (ten) years.
Radiation surveys are recommended to be conducted on a regular base in accordance with
local policy.
In case of any device malfunctions related to the radiation source the manufacturer is to be
informed immediately.