Freiberg
Instruments
GmbH
ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
63 / 78
operator near the device
2.2.2 Exchange of optical detector/adjustment of optical components
Process steps
1. Opening of the
lexsyg
housing
2. Disassembling of the optical system mounted on
top of the radiation source module
3. Assembling
and
adjusting
the
optical
components
4. Test run
5. Closing the
lexsyg
housing
duration
20 min
Typical distance to source
30 cm (eyes, body)
10 cm (hands)
Dose per handling
13 µSv (eyes, body)
55 µSv (hands)
Number of annual activities
20
notes
Shutter typically closed / can be open for a short
period of time
Beta radiation completely shielded by radiation
source module
Housing of the
lexsyg
opened
If applicable, use appropriate means of radiation
protection (tools)