![Freiberg Instruments lexsyg research Скачать руководство пользователя страница 60](http://html1.mh-extra.com/html/freiberg-instruments/lexsyg-research/lexsyg-research_manual-and-documentation_2331334060.webp)
Freiberg
Instruments
GmbH
ּ Am St. Niclas Schacht 13 ּ 09599 Freiberg – Germany
60 / 78
The radiation source is precisely positioned and safely fixed in the source holder by three
special screws (see fig. 6, right) ensuring that the radiation exit window cannot be touched or
even damaged by the shutter.
The following information is laser engraved on the circumference of the source body (see
fig. 8):
-
Radioactive trefoil symbol
-
Radioisotope: Sr-90
-
Maximum activity: 2 GBq
-
Serial number: RFQ-xxx (xxx may range from 001 to 999)
Further, the radiation source module is uniquely identified by type and serial number (see fig.
4). The labelling of the source module also provides information on the radiation source
installed. The housing bears the radiation trefoil symbol.
Table 1
Technical specifications of the single miniaturized sources the circular source
consists of (refer to attachment 1)
Source type
strontium-90 seed type Sr0.S03-1
manufacturer
Eckert & Ziegler BEBIG GmbH
radioisotope
Sr-90
maximum activity
118 MBq
radioactive material
physical
solid
chemical
SrTiO
3
(strontium titanate)
form
sealed
source canister
stainless steel 1.4301/1.4541, 40 µm thick, laser welded
dimensions
length: 2.5 mm, diameter: 0.64 mm
type of radiation
beta
maximum energy
2.27 MeV
ISO 2919 classification
C.63211
special tests
ANSI N44.1-1977 tests: puncture, impact, temperature
leakage test
immersion test per ISO 9978