48 / 78
Now you have free access to the detection window. You can remove it by opening the tree
screws in the frame to clean or replace it.
figure 31:
overview of the OSL-unit
Mounting works in reverse direction. Don’t forget to mount the two gaskets below the window
and into the window frame.
Cleaning at the extra TL position (optional):
First remove the plug and the air tube backside of the filter changing unit on this position. For
removing the air tube push on the small frame of the connector and pull the air tube out.
figure 32:
removing the plugs of the extra TL position
Open the small fastener by pulling and turning. (see below)
figure 33:
fastener of the extra Tl position
Now open whole unit by pulling it up. It is blocked in straight position by himself. (see below)