Module 5.1.1: UV-VIS PMT Module
for standard applications in luminescence dating and dosimetry
detection unit with standard bialkaline cathode PMT (HAMAMATSU)
sensitivity 280-650 nm
peak sensitivity 420 nm ( 3.1 E5 counts s
-1
pW
-1
)
dark counts typ. 100 cps
photon counter
counting linearity (random pulses, 10% loss) 6.0 E6 cps
electromechanical shutter, UV-VIS optics
Module 5.1.2: Extended UV-Vis PMT Module
for standard applications in luminescence dating and dosimetry
detection unit with standard bialkaline cathode PMT (Electron Tubes 9235Q)
sensitivity 160-630 nm
peak sensitivity: 30% quantum efficiency at 200 nm and 350 nm
dark counts typ. 300 cps
photon counter
counting linearity (random pulses, 10% loss): must be specified
electromechanical shutter, UV-VIS optics
Module 5.1.3: Red enhanced UV-Vis PMT Module
necessary/useful if standard UV-VIS PMT-modules yellow to red sensitivity is not
sufficient
useful e.g. for quartz(flint)/calcite orange-red or feldspar red luminescence
measurements
thermoelectric/air cooled GaAsP photocathode PMT (HAMAMATSU)
sensitivity: 300
– 720 nm,
peak-sensitivity: 40% quantum efficiency at 580 nm
cooling max. 35 deg. below RT
cooling time approx. 5 min
typ. 100 cps dark signal at 0°C
photon counting system
counting linearity (random pulses, 10% count loss) 1.5 E6 cps