The
lexsyg
measurement system provides space for up to 80 sample discs. The actual
allocation can be determined at the sample list. It provides a list with an entry for every
sample slot containing the
name
of the containing sample, the
slot
- number, the associated
sequence
and the
status
of the sample slot.
Warning:
For mechanical reasons never use a magnetic sample disc at position 21!
Otherwise it is possible to lose this disc during the sample transport.
To load (or unload) a sample disc, the sample wheel must be rotated until the required slot is
in front of the sample storage. There a two ways to achieve this. First, it is possible to select
the relevant slot in the numeric up/down
move position to Load/Unload
and press the
button
move
. The wheel will now move the sample slot in front of the sample storage. The
second way is to go to the front side of the sample storage and press the
sample wheel
forward/reverse
buttons (see Figure 2) until the required slot is in front. Now it is possible to
add or remove the sample disc to or from the slot.
After this, the new status of the slots must be announced to the software. Do this by selecting
the slot in the
sample list
and change the
status of the marked positions
to one of the
three possibilities:
empty
– the slot contains no sample disc and so will not be processed.
loaded
– the slot contains a sample disc and will be processed when a sequence is
associated and the measurement system is running.
measured - the slot contains a sample disc but will not be processed by the
measurement system. After a measurement the system will change the status of a
slot to this state.
Be sure to always select the correct new state of a sample slot after loading/ unloading a
sample. Otherwise malfunctions of the system are possible (e.g. unloading in an already
occupied slot).
When the sample discs are loaded, a measurement sequence can be created and linked to a
sample slot.
You can select more than one slot at a time for an operation by using the
strg
and/ or
shift
key while clicking in the sample list.