2.2.3 Exchange of quartz glass window (top of the radiation source module)
Process steps
1. Opening of the
lexsyg
housing
2. Disassembling of the external optical system
mounted on top of the radiation source module
3. Removing the lead shield of the radiation source
module
4. Use of a temporary radiation shield (10 mm
acrylic glass wall)
5. Removing the window fixing ring
6. Exchanging the window (use tweezers or similar
tools)
7. Assembling the window fixing ring
8. Assembling
and
adjusting
the
optical
components
9. Test run
10. Closing the
lexsyg
housing
duration
20 min
Typical distance to source
30 cm (eye, body)
10 cm (hands)
Dose per handling
13 µSv (eye, body)
55 µSv (hands)
Number of annual activities
1
notes
Shutter typically closed / can be open for a short
period of time
Until window is removed beta radiation is completely
shielded by radiation source module
Upon removing the quartz window beta radiation
may escape from the radiation source module
For step 6 use tweezers or other appropriate tool;
wear radiation protection goggles