60
Chapter 2 Working with DaqLab
Figure 19: Two-point calibration dialog box
Note:
We strongly recommend calibrating using the values that appear in the two Reference value
text boxes
e. If not selected, click the
Calibration
tab.
f. Click
Restore defaults
, then click
OK.
g. Measure two known reference values and record both the reference values and
the values displayed by DaqLab.
Enter calibration values
a. Repeat steps 1 to 5 above.
b. Select
the
Two-point calibration
option.
c. Enter a distinct reference value in each of the
Reference Value
text boxes and the
corresponding logger values in each of the
Logger Value
text boxes (the logger
values are the values displayed by DaqLab when measuring the two reference
values).
d. Click
OK.