Calibration Adjustment
5.6 Final Calibration
5
5-15
al55ex2w.bmp
Figure 5-7. 5500A NORMAL Output to Test Tool Banana Input
3.
Set the 5500A to supply a DC voltage, to the first calibration point in Table 5-5.
4.
Set the 5500A to operate (OPR).
5.
Press
to start the calibration.
6.
Wait until the display shows calibration status
:READY
.
7.
Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points of Table 5-4
8.
Set the 5500A to Standby, and continue at Section 5.6.7.
Table 5-5. DMM Gain Calibration Points
Cal step
UUT input value (5500A NORMAL)
CL 0840
500 mV
CL 0849
2.5 V
CL 0841
5 V
CL 0842
50 V (set 5500A to OPR!)
CL 0843
500 V
CL 0844
1000 V
5.6.7 Input A& B, and DMM Zero
Proceed as follows to do the Input A&B, and the DMM Zero calibration:
1.
Press
to select calibration adjustment step CL0852
2.
Short circuit Input A and Input B.
Содержание ScopeMeter 19 B Series