
Performance Verification
4.6 Scope Input A&B Tests
4
4-25
Proceed as follows:
1.
Connect the test tool to the calibrator as shown in Figure 4-16.
al55sca.bmp
Figure 4-16. Test Tool Input A to TV Signal Generator
2.
Select the following test tool setup:
•
Reset the test tool
•
Press
, then press
to open the Trigger Options menu.
•
Choose
VIDEO on A...
, then from the shown opened menu choose
Polarity: POSITIVE | PAL
(
or
NTSC PALplus SECAM
)
•
Press
to select
ALL LINES
•
Press
to enable the arrow keys for selecting the video line number.
•
Using
select line number:
⇒
622 for PAL, PALplus, or SECAM
⇒
525 for NTSC.
•
Using
and
set the Input A sensitivity to 2 V/div (the actual probe setting
is 10:1).
•
Using
select the time base to 20
μ
s/div.
3.
Set the calibrator to mode video with amp100%. Set format and marker line
number to :
⇒
PAL 622 (even), for PAL and PALplus
⇒
SECAM 622 (even), for SECAM
⇒
NTSC 262 even, for NTSC.
4.
Observe the trace, and check to see if the test tool triggers on the negative pulse
before the marker pulse (see Figure 17).
Содержание ScopeMeter 19 B Series
Страница 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Страница 8: ......
Страница 10: ......
Страница 14: ...2 2 ...
Страница 16: ......
Страница 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Страница 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Страница 58: ......
Страница 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Страница 78: ......
Страница 86: ...Fluke 19xB 19xC 2x5C Service Manual 6 10 ...