Fluke 19xB-19xC-2x5C
Service Manual
4-20
4.6.11 Input B Trigger Sensitivity Test
Proceed as follows to test the Input B trigger sensitivity:
1.
Connect the test tool to the 5500A as shown in Figure 4-7.
al55scb.bmp
Figure 4-7. 5500A Scope Output to Test Tool Input B
2.
Select the following test tool setup:
•
Reset the test tool
•
Press
and use
to turn Input B on.
•
Press
and use
to turn Input A off.
•
Using
move the Input B trace zero to the center grid line.
•
Press
and use
to select Input B as trigger source.
•
Using
and
change the sensitivity range to select manual sensitivity
ranging, and lock the Input B sensitivity range on 2 V/div.
3.
Using
select the time base indicated under the first column of Table 4-10.
4.
Set the 5500A to source the leveled sine wave given in the first row of Table 4-10.
5.
Adjust the 5500A output voltage until the displayed trace has the amplitude indicated
under the appropriate column of Table 4-10.
6.
Verify that the signal is well triggered.
If it is not, press
, then using
enable the up/down arrow keys for manual
Trigger Level adjustment. Adjust the trigger level and verify that the signal will be
triggered now. The trigger level is indicated by the trigger icon ( ).
7.
Continue through the test points.
8.
When you are finished, set the 5500A to Standby.
Содержание ScopeMeter 19 B Series
Страница 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Страница 8: ......
Страница 10: ......
Страница 14: ...2 2 ...
Страница 16: ......
Страница 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Страница 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Страница 58: ......
Страница 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Страница 78: ......
Страница 86: ...Fluke 19xB 19xC 2x5C Service Manual 6 10 ...