Fluke 19xB-19xC-2x5C
Service Manual
4-18
•
Using
and
change the sensitivity range to select manual sensitivity
ranging, and lock the Input A sensitivity range on 2 V/div.
3.
Using
select the time base indicated under the second column of Table 4-8.
4.
Set the 5500A to source the leveled sine wave for the appropriate test tool model.
5.
Adjust the 5500A output voltage until the displayed trace has the trigger amplitude
indicated under the last column of Table 4-8.
6.
Verify that the signal is well triggered.
If it is not, press
, then using
enable the up/down arrow keys for manual
Trigger Level adjustment. Adjust the trigger level and verify that the signal will be
triggered now. The trigger level is indicated by the trigger icon ( ).
7.
Continue through the test points.
8.
When you are finished, set the 5500A to Standby.
Table 4-8. Input A Trigger Sensitivity Test Points
UUT
UUT
5500A SC... MODE levsin
UUT
Model
Time base
Initial Input Voltage Frequency Trigger
Amplitude
ALL
200 ns/div
100 mV pp
5 MHz
0.5 div
192B-C
10 ns/div
400 mV pp
60 MHz
1 div
10 ns/div
800 mV pp
100 MHz
2 div
196B-C, 215C
10 ns/div
400 mV pp
100 MHz
1 div
10 ns/div
800 mV pp
150 MHz
2 div
199B-C, 225C
10 ns/div
400 mV pp
200 MHz
1 div
10 ns/div
800 mV pp
250 MHz
2 div
Содержание ScopeMeter 19 B Series
Страница 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Страница 8: ......
Страница 10: ......
Страница 14: ...2 2 ...
Страница 16: ......
Страница 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Страница 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Страница 58: ......
Страница 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Страница 78: ......
Страница 86: ...Fluke 19xB 19xC 2x5C Service Manual 6 10 ...