Fluke 19xB-19xC-2x5C
Service Manual
5-12
Table 5-3. Input A&B Gain Calibration Points
Cal step
UUT input value (5500A SCOPE, MODE volt, SCOPE Z 1 M
Ω
, 1 kHz)
CL 0660
300 mV
CL 0604
500 mV
CL 0637
none
(5500 standby)
CL 0504
500 mV
CL 0624
none
(5500 standby)
CL 0599
Not for software versions
V05.01 and V05.02
10
mV
CL 0600
25 mV
CL 0601
50 mV
CL 0602
100 mV
CL 0603
250 mV
CL0662
2
V
CL 0605
1 V
CL 0606
2.5 V
CL 0607
5 V
CL0664
20
V
CL 0608
10 V
CL 0609
25 V
CL 0610
50 V
(set 5500A to OPR!)
5.6.4 Input A&B Position
Proceed as follows to do the Input A&B Position calibration:
1.
Press
to select calibration adjustment step
CL 0620
(software versions V05.01
and V05.02), or
CL 0619
(software versions V05.03 and newer).
2.
Remove all Input A and Input B connections (open inputs).
3.
Press
to start the calibration
4.
Wait until the display shows calibration status
:READY
.
5.
Continue at Section 5.6.5
Содержание ScopeMeter 19 B Series
Страница 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Страница 8: ......
Страница 10: ......
Страница 14: ...2 2 ...
Страница 16: ......
Страница 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Страница 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Страница 58: ......
Страница 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Страница 78: ......
Страница 86: ...Fluke 19xB 19xC 2x5C Service Manual 6 10 ...