Fluke 19xB-19xC-2x5C
Service Manual
5-10
al55scb.bmp
Figure 5-4. 5500A SCOPE Output to Test Tool Input B
3.
Set the 5500A SCOPE output to source the signal required for the first calibration
point in Table 5-2.
4.
Set the 5500A in operate (OPR) or standby (STBY) as indicated.
5.
Press
to start the calibration.
6.
Wait until the display shows calibration status
:READY
.
7.
Press
to select the next calibration step, set the 5500A to the next calibration
point signal, and start the calibration.
Continue through all calibration points of Table 5-2.
8.
When you are finished, set the 5500A to Standby.
9.
Continue at Section 5.6.3.
Table 5-2. Input B LF-HF Gain Calibration Points
Cal step
UUT input signal
5500A Setting
CL 0674
none
STANDBY
CL 0410
0.5 Vpp square wave, 1 kHz
SCOPE edge, 0.5 Vpp, 1 kHz
CL 0724
none
STANDBY
CL 0421
0.5 Vpp square wave, 1 kHz
SCOPE edge, 0.5 Vpp, 1 kHz
CL 0482
0.5 Vpp sine wave, 50 kHz
SCOPE levsine, 0.5 Vpp, 50 kHz
CL 0483
0.5 Vpp sine wave
Fluke 199B-C, 225C: 221 MHz
Fluke 196B-C, 215C: 141 MHz
Fluke 192B-C:
91 MHz
SCOPE levsine, 0.5 Vpp,
221
MHz
141
MHz
91
MHz
Содержание ScopeMeter 19 B Series
Страница 6: ...Fluke 19xB 19xC 2x5C Service Manual iv ...
Страница 8: ......
Страница 10: ......
Страница 14: ...2 2 ...
Страница 16: ......
Страница 24: ...Fluke 19xB 19xC 2x5C Service Manual 3 10 ...
Страница 26: ...Fluke 19xB 19xC 2x5C Service Manual 4 2 ...
Страница 58: ......
Страница 76: ...Fluke 19xB 19xC 2x5C Service Manual 5 20 ...
Страница 78: ......
Страница 86: ...Fluke 19xB 19xC 2x5C Service Manual 6 10 ...