5820A
Service Manual
3-36
Connect the Calibrator Mainframe CHAN 1 connector to the 5790A WIDEBAND input
with the BNC(f) to Type N(m) adapter. Set the 5790A to AUTORANGE, digital filter
mode to FAST, restart fine, and Hi Res on.
3-45. Equipment Setup for High Frequency Flatness
All high frequency flatness procedures use the following equipment:
•
Hewlett-Packard 437B Power Meter
•
Hewlett-Packard 8482A and 8481D Power Sensors
•
BNC(f) to Type N(f) adapter
•
N to BNC cable supplied with the Calibrator Mainframe
Note
When high frequencies at voltages below 63 mV p-p are verified, use the
8481D Power Sensor. Otherwise, use the 8482A Power Sensor.
Connect the HP 437B Power Meter to either the 8482A or the 8481D Power Sensor. For
more information on connecting the two instruments, see the power meter and power
sensor operators manuals.
Connect the power meter/power sensor combination to the CHAN 1 connector on the
Calibrator Mainframe.
The Hewlett-Packard 437B Power Meter must be configured by setting the parameters
listed below. Zero and self-calibrate the power meter with the power sensor being used.
Refer to the Hewlett-Packard 437B Operators Manual for details.
•
PRESET
•
RESOLN 3
•
AUTO FILTER
•
WATTS
•
SENSOR TABLE 0 (default)
3-46. Low Frequency Verification
This procedure provides an example of testing low frequency flatness using a 5.5 V
output. Follow the same procedure for testing other amplitudes, only compare results
against the flatness specification listed in Table 3-20.
1.
Program the Calibrator Mainframe for an output of 5.5 V @ 500 kHz. Press
on
the Calibrator Mainframe to activate the output.
2.
Allow the 5790A reading to stabilize. The 5790A should display approximately 1.94
V rms. Enter the 5790A reading in Column A of Table 3-20.
3.
Enter 50 kHz into the Calibrator Mainframe. Allow the 5790A reading to stabilize,
then enter the 5790A reading in Column B of Table 3-20.
4.
Enter the next frequency listed in Table 3-20. Allow the 5790A reading to stabilize,
then enter the reading into Column A of the table.
5.
Enter 50 kHz into the Calibrator Mainframe. Allow the 5790A reading to stabilize,
then enter the 5790A reading in Column B of Table 3-20.
6.
Repeat steps 4 and 5 for all of frequencies listed in Table 3-20. Continue until you
have completed Columns A and B.
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...