Introduction and Specifications
General Specifications
1
1-15
1-23. Overload Measurement Specifications
The Overload test function applies dc or ac (1 kHz square wave) power into the 50
Ω
oscilloscope input and monitors the current. A time measurement counter indicates the
time duration of the applied overload signal. When the input protection circuit reacts and
opens up the 50
Ω
load, the calibrator indication is set to ‘off’ on the right hand display.
In order to prevent front end damage to the oscilloscope, a limited amount of energy is
applied by a user selectable time limit.
Table 1-17. Overload Measurement Specifications
Source
Voltage
Typical ‘On’ current
indication
Typical ‘Off’ current
indication
Maximum Time Limit DC or
AC (1 kHz)
5 V to 9 V
100 mA to 180 mA
10 mA
setable 5 to 60 sec
1-24. External Reference Input Specifications
The External Reference Input selection allows the user to provide their own high
stability 10 MHz reference clock for the 5820A for all functions except the Wave
Generator function. For all other modes, the frequency stability is determined by the
external reference stability. The external reference input must be between 1 to 5 V p-p.
1-25. Auxiliary Input Specifications
Maximum input voltage into the auxiliary input is 40 V p-p.
Table 1-18. Auxiliary Input Performance
Channel Configuration
Frequency
Typical Loss
Typical VSWR
1-Channel
< 600 MHz
≤
1.1 dB
≤
1.2:1
1-Channel
600 MHz to 1 GHz
≤
1.3 dB
≤
1.4:1
1-Channel
1 GHz to 2.0 GHz
≤
2.0 dB
≤
1.7:1
1-Channel
2 GHz to 3 GHz
≤
3.0 dB
≤
2.0:1
5-Channel
< 600 MHz
≤
1.1 dB
≤
1.2:1
5-Channel
600 MHz to 1 GHz
≤
1.3 dB
≤
1.4:1
5-Channel
1 GHz to 2.0 GHz
≤
2.0 dB
≤
1.7:1
5-Channel
2 GHz to 3 GHz
≤
3.0 dB
≤
2.0:1
1-26. Current Output Specifications
Table 1-19. Current Output Specifications
DC
Squarewave
Amplitude (compliance voltage 2 V max)
±
100 µA to
±
100 mA
100 µAp-p to 100 mAp-p
Accuracy
±
(0.25% + 0.5 µA)
±
(0.25% + 0.5 µA) [1]
Frequency Range
N/A
10 Hz to 100 KHz
Accuracy
2.5 ppm +5 µHz
Steps
1,2,5 or continuous
[1] Amplitude uncertainty for frequency range 45 Hz to 1 kHz at < 120 mV compliance voltage.
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...