5820A
Service Manual
3-62
•
Output cable provided with the 5820A
Before you begin this procedure, verify that the 5820A is in the edge mode (the Edge
menu is displayed), and program it to output 1 V p-p @ 1 MHz. Press
to activate
the output.
Refer to Figure 3-6 for the proper setup connections and connect the Calibrator
Mainframe to the oscilloscope. Set the oscilloscope vertical to 10 mV/div and horizontal
to 1 ns/div. Set the oscilloscope to look at the 90% point of the edge signal; use this
point as the reference level. Set the oscilloscope to look at the first 10 ns of the edge
signal with the rising edge at the left edge of the oscilloscope display.
3-69. Adjusting the Edge Aberrations
Refer to Figure 3-12 while making the following adjustments:
1.
Adjust A90R13 to set the edge signal at the right edge of oscilloscope display, at 10
ns, to the reference level set above.
2.
Adjust A90R36 so the first overshoot is the same amplitude as the next highest
aberration.
3.
Adjust A90R35 so that the second and third overshoot aberrations are the same
amplitude as the first aberration.
4.
Adjust A90R12 to set the edge signal occurring between 2 ns and 10 ns to the
reference level set above.
5.
Readjust A90R36 and A90R35 to obtain equal amplitudes for the first, second, and
third aberrations.
6.
Adjust A90R13 to set the edge signal occurring between 0 ns and 2 ns to the
reference point set above. Center any aberrations so the peaks are equal above and
below the reference level.
7.
Readjust A90R12 if necessary to keep the edge signal occurring between 2 ns and 10
ns at the reference level.
8.
Readjust A90R13 if necessary to keep the edge signal occurring between 0 ns and 2
ns at the reference level.
9.
Set the UUT output to 250 mV and the oscilloscope vertical to 2 mV/div. Check the
aberrations.
10.
Connect the 10 dB attenuator to the oscilloscope input. Connect the UUT to the
attenuator and program the UUT output to 2.5 V.
11.
Set the oscilloscope vertical to 5 mV/div. Check the aberrations.
12.
Check for rise time < 300 ps at 250 mV, 1 V, and 2.5 V outputs.
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...