5820A
Service Manual
3-12
5.
Plug in the 100
Ω
precision resistor into the Calibrator front panel.
6.
Take voltage measurement across the resistor at 9.00 mA, convert to mA, and enter
the DMM value on the Calibrator front panel.
7.
Press the GO ON blue softkey.
8.
Plug in the 10
Ω
precision resistor into the Calibrator front panel.
9.
Take voltage measurement across the resistor at 90.0 mA, convert to mA, and enter
the DMM value on the Calibrator front panel.
10.
Press the GO ON blue softkey.
11.
To complete the Current calibration, press the STORE CONSTS blue softkey to
store the new calibration constants.
Note
The Calibrator Mainframe will warn when the entered value is out of
bounds. If this warning occurs, check the setup again and carefully re-
enter the reading insuring proper multiplier (i.e., m,
µ
, n, p). If the warning
still occurs, repair may be necessary.
3-14. Wave Generator Calibration
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter
•
BNC(f) to Double Banana adapter
•
N to BNC cable supplied with the Scope Calibrator
Within the calibration menu, press the NEXT SECTION blue softkey until the display
reads “WAVEGEN Cal:”. Then follow these steps to calibrate the Wave Generator:
1.
Connect the Calibrator Mainframe’s CHAN 1 connector to the HP 3458A input,
using the N to BNC cable and the BNC(f) to Double Banana adapter.
2.
Set the HP 3458A to DCV, NPLC = .01, LEVEL 1, TRIG LEVEL, and the DELAY
to .0002 for measuring the upper part of the wave form (i.e. topline), and the
DELAY to .0007 for measuring the lower part of the wave form (i.e. baseline).
Manually range lock the HP 3458A to the range that gives the most resolution for
the topline measurements. Use this same range for the corresponding baseline
measurements at each step.
3.
For each calibration step, take samples for at least two seconds, using the HP 3458A
MATH functions to retrieve the average or mean value. See “Setup for Scope
Calibrator Edge and Wave Generator Measurements” for more details.
3-15. Edge Amplitude Calibration
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter
•
BNC(f) to Double Banana adapter
•
N to BNC cable supplied with the Scope Calibrator
•
50
Ω
feedthrough termination
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...