5820A
Service Manual
3-18
Note
Calibrating the capacitance for the Scope Calibrator uses 4 capacitors,
0 pF, 5 pF, 28 pF, and 50 pF. Starting with 0 (or no input) you proceed
through each capacitor entering the exact value each time.
12.
When prompted for the first reference capacitor by the Calibrator Mainframe,
disconnect the 1.5 M
Ω
resistance and leave nothing attached to the end of the N to
BNC cable.
13.
Press the GO ON blue softkey.
14.
Enter 0.
15.
When prompted by the Calibrator Mainframe, connect the 5 pF capacitor to the end
of the N to BNC cable.
16.
Press the GO ON blue softkey.
17.
Enter the actual 5 pF capacitance value.
18.
When prompted by the Calibrator Mainframe, disconnect the 5 pF capacitor and
connect the 28 pF to the end of the N to BNC cable.
19.
Press the GO ON blue softkey.
20.
Enter the actual 28 pF capacitance value.
21.
When prompted by the Calibrator Mainframe, disconnect the 28 pF capacitor and
connect the 50 pF to the end of the N to BNC cable.
22.
Press the GO ON blue softkey.
23.
Enter the actual 50 pF capacitance.
24.
The Calibrator Mainframe will prompt that the calibration is complete. Press the
STORE CONSTS blue softkey to store the new calibration constants.
3-22. Leveled Sine Wave Flatness Calibration (GHz Option)
Leveled Sine Wave Flatness Calibration is only applicable if the 5820A has the GHz
Option installed. This calibration procedure uses the same equipment as in the 5820A
Leveled Sine Wave Flatness (High Frequency) calibration procedure. See the GHz
section for details.
3-23. 5820A-5 Option
The 5820A-5 Option allows you to calibrate up to five oscilloscope channels
simultaneously without changing cables. This allows you to perform fast, automated
calibrations with documented procedures and results while freeing the operator to
complete other work. You can find this option discussed throughout the manual where
appropriate.
Note
If the 5820A is equipped with the 5-channel option, the Mainframe will
indicate when to move to the next channel.
3-24. Verification
All of the Oscilloscope Calibration functions should be verified at least once per year, or
each time the Scope Calibrator is calibrated. The verification procedures in this section
provide traceable results; however the factory uses different procedures of higher
precision than those described here. The procedures in this manual have been developed
to provide users the ability to verify the Scope Calibrator at their own site if they are
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...