Calibration and Verification
DC Voltage Calibration
3
3-9
CHAN 2
CHAN 4
EXT TRIG
EXT TRIG
SOURCE/MEASURE
CHAN 1
CHAN 2
CHAN 3
CHAN 4
CHAN 5
20V PK
MAX
CHAN 1-5
AUX
INPUT
20V PK
MAX
20V PK
MAX
CAT
130V
PK
MAX
SOURCE
30V DC
MAX
MEASURE
5820A
OSCILLOSCOPE
CALIBRATOR • 2GHz
5820A
HP 3458A
5820A Cable
50
Ω
Feedthrough
Termination
BNC(F) to
Double Banana
Adapter
yu055f.eps
Figure 3-2. Setup for Scope Calibrator Edge and Wave Gen Square Wave Measurements
For all measurements, the HP 3458A is in DCV, manual ranging, with level triggering
enabled. A convenient method to make these measurements from the HP 3458A’s front
panel is to program these settings into several of the user defined keys on its front panel.
For example, to make topline measurements at 1 kHz, you would set the DMM to
“NPLC .01; LEVEL 1; DELAY .0002; TRIG LEVEL”. To find the average of multiple
readings, you can program one of the keys to “MATH OFF; MATH STAT” and then use
the “RMATH MEAN” function to recall the average or mean value.
3-8. DC Voltage Calibration
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter
•
BNC(f) to Double Banana adapter
•
N to BNC cable supplied with the Scope Calibrator
Note
Calibrating dc voltage requires ac voltage calibration.
Note
When dc voltage calibration is started, the calibrator uses a special
configuration of WAVEGEN mode (with a zero frequency, ordinarily
undefined) to source approximately 3 Vdc. After the operator enters the
true dc value, this value is then transferred to the internal DACs and A/D
through a series of internal calibration steps. On firmware version 1.5 and
earlier, the output displayed is 6 V-pp @ 0.00 Hz at this step. This is a
side-effect of the special configuration used. As of firmware version 1.6,
the output display indicates 3 V @ 0.00 Hz.
Set the Calibrator to Cal DCV by pressing the SETUP, CAL and 5820A CAL blue
softkeys then follow these steps to calibrate dc voltage.
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...