5820A
Service Manual
3-14
Note
The Calibrator Mainframe will warn when the entered value is out of
bounds. If this warning occurs recheck the setup and calculation and
carefully re-enter the corrected rms reading insuring proper multiplier
(i.e., m, u, n, p). If the warning still occurs, repair may be necessary.
6.
Repeat step 5 until the Calibrator Mainframe display indicates that the next steps
calibrate Leveled Sine flatness. Press the STORE CONSTS blue softkey to store the
new calibration constants.
I
O
0
•
1
2
3
4
5
6
7
8
9
ENTER
k
sec
m
µ
V
NEW
REF
Hz
FIELD
EDIT
/
+
VOLT
EDGE
LEVSINE
MARKER
MENU
PREV
M
RESET
SETUP
CE
MODES
MORE
CHAN
dBm
n
MULT
x
DIV
÷
INPUT
AUX
OPR
STBY
CHAN 2
CHAN 4
EXT TRIG
EXT TRIG
SOURCE/MEASURE
CHAN 1
CHAN 2
CHAN 3
CHAN 4
CHAN 5
20V PK
MAX
CHAN 1-5
AUX
INPUT
20V PK
MAX
20V PK
MAX
CAT
130V
PK
MAX
SOURCE
30V DC
MAX
MEASURE
5820A
OSCILLOSCOPE
CALIBRATOR • 2GHz
POWER
I
O
5790A
AC MEASUREMENT
STANDARD
1000V RMS MAX
1000V RMS MAX
3V RMS MAX
INPUT 1
INPUT 2
SHUNT
HI
LO
GUARD
GROUND
SHELL FLOATING
10V PEAK
MAX
WIDEBAND
7V RMS MAX
MAX
10V PK
SHELL FLOATING
INPUT1
SHUNT
INPUT1 INPUT1
INPUT1
2.2 mV
22 mV
220 mV
7 mV
70 mV
700 mV
2.2 V
22 V
220 mV
7V 70V
700V
1kV
2.2 mV
AUTO
MAN
VIEW
REF
UTIL
MENUS
SPEC
CLEAR
DELETE
ENTER
0
1
2
3
4
5
6
7
8
9
+/-
.
yu034f.eps
Figure 3-3. Connecting the Calibrator Mainframe to the 5790A AC Measurement Standard
3-17. Leveled Sine Wave Flatness Calibration
Leveled Sine Wave flatness calibration is divided into two frequency bands: 50 kHz to
10 MHz (low frequency) and >10 MHz to 600 MHz (high frequency). The equipment
setups are different for each band. Flatness calibration of the low frequency band is
made relative to 50 kHz. Flatness calibration of the high frequency band is made relative
to 10 MHz.
Leveled Sine Wave flatness is calibrated at multiple amplitudes. Both low and high
frequency bands are calibrated at each amplitude. Calibration begins with the low
frequency band, then the high frequency band for the first amplitude, followed by the
low frequency band, then the high frequency band for the second amplitude, and so on,
until the flatness calibration is complete.
Press the NEXT SECTION blue softkey until the display reads “Set up to measure
leveled sine flatness”.
3-18. Low Frequency Calibration
Connect the Calibrator Mainframe CHAN 1 connector to the 5790A WIDEBAND input
as described under “Equipment Setup for Low Frequency Flatness”.
Follow these steps to calibrate low frequency Leveled Sine Wave flatness for the
amplitude being calibrated.
1.
Press the GO ON blue softkey.
2.
Establish the 50 kHz reference:
•
Allow the 5790A rms reading to stabilize.
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...