![Fluke 5820A Скачать руководство пользователя страница 42](http://html1.mh-extra.com/html/fluke/5820a/5820a_service-manual_2309653042.webp)
5820A
Service Manual
3-4
Table 3-1. Scope Calibrator Calibration and Verification Equipment (cont.)
Current Calibration, AC/DC Current Verification Minimum Use Specifications
Instrument
Model
Digital
Multimeter
HP 3458A
Voltage
1 mV to
±
2 V p-p Uncertainty: 0.06%
Adapter
Pomona #1269
BNC(f) to Double Banana Plug
1
Ω
Resistor
Assembly
User Supplied
Measured to
±
0.05%, 125 mW, including banana jack
mounting
10
Ω
Resistor
Assembly
User Supplied
Measured to
±
0.05%, 250 mW, including banana jack
mounting
100
Ω
Resistor
Assembly
User Supplied
Measured to
±
0.05%, 125 mW, including banana jack
mounting
1000
Ω
Resistor
Assembly
User Supplied
Measured to
±
0.05%, 125 mW, including banana jack
mounting
N to BNC
Cable
(supplied with Scope
Calibrator)
Pulse Width Calibration and Verification
High-Frequency Digital
Storage Oscilloscope
Tektronix 11801 with Tektronix SD-
22/26 sampling head or Tektronics
TDS 820 scope with 8 GHz option.
Attenuator (2)
3 dB, SMA (m/f)
SMA T connector
(m)-(f)-(m)
0.5 M SMA cable
(m)-(m)
Adapter (2)
BNC(f) to SMA(m)
N to BNC Cable (2)
(supplied with Scope Calibrator)
Leveled Sine Wave Frequency Verification
Frequency
Counter
PM 6680 with option (PM 9621, PM 9624,
or PM 9625) and (PM 9690 or PM 9691)
50 kHz to 600 MHz, <0.1 ppm
uncertainty
Adapter
Pomona #3288
BNC(f) to Type N(m)
N to BNC Cable
(supplied with Scope Calibrator)
Leveled Sine Wave Flatness (Low Frequency) Calibration and Verification
AC Measurement
Fluke 5790A
Range
5 mV p-p to 5.5 V p-p
Standard
with -03 option
Frequency
50 kHz to 10 MHz
Adapter
Pomona #3288
BNC(f) to Type N(m)
N to BNC Cable
(supplied with Scope
Calibrator)
Leveled Sine Wave Harmonics Verification
Spectrum Analyzer
HP 8590A
600 MHz and below
HP 8592L
Above 600 MHz (GHz Option)
Adapter
Pomona #3288
BNC(f) to Type N(m)
N to BNC Cable
(supplied with Scope
Calibrator)
Pulse Period, Edge Frequency, AC Voltage Frequency Verification
Frequency Counter
PM 6680 with option (PM
9690 or PM 9691)
20 ms to 150 ns, 10 Hz to 10 MHz: < 0.12 ppm
uncertainty
N to BNC Cable
(supplied with Scope
Calibrator)
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...