Calibration and Verification
Pulse Width Calibration
3
3-15
•
Press the 5790A Set Ref blue softkey. (Clear any previous reference by pressing
the 5790A Clear Ref blue softkey prior to setting the new reference if required.)
3.
Press the GO ON blue softkey.
4.
Adjust the amplitude using the Calibrator Mainframe front panel knob until the
5790A reference deviation matches the 50 kHz reference within 1000 ppm.
5.
Repeat steps 1 to 4 until the Calibrator Mainframe display indicates that the
reference frequency is now 10 MHz. Continue with the high frequency calibration.
3-19. High Frequency Calibration
Connect the Calibrator Mainframe CHAN 1 connector to the power meter and power
sensor as described under “Equipment Setup for High Frequency Flatness”.
Follow these steps to calibrate high frequency Leveled Sine Wave flatness for the
amplitude being calibrated.
1.
Press the GO ON blue softkey.
2.
Establish the 10 MHz reference:
•
Press the power meter SHIFT key, then FREQ key and use the arrow keys to
enter the power sensor’s 10 MHz Cal Factor. Ensure that the factor is correct,
then press the power meter ENTER key.
•
Allow the power meter reading to stabilize.
•
Press the Power meter REL key.
3.
Press the GO ON blue softkey.
4.
Press the power meter SHIFT key, then FREQ key and use the arrow keys to enter
the power sensor’s Cal Factor for the frequency displayed on the Calibrator
Mainframe. Ensure that the factor is correct, then press the power meter ENTER
key.
5.
Adjust the amplitude using the Calibrator Mainframe front panel knob until the
power sensor reading matches the 10 MHz reference within 0.1%.
6.
Repeat steps 1 to 5 until the Calibrator Mainframe display indicates that either the
reference frequency is now 50 kHz or that the next steps calibrate pulse width.
Repeat the low frequency calibration procedure for the next amplitude unless the
Calibrator Mainframe display indicates that the next steps calibrate pulse width.
Press the STORE CONSTS blue softkey to store the new calibration constants.
3-20. Pulse Width Calibration
This procedure uses the following equipment:
•
High Frequency Digital Storage Oscilloscope: Tektronix 11801 with Tektronix SD-
22/26 sampling head
•
3 dB attenuator, 3.5 mm (m/f)
•
BNC(f) to 3.5 mm(m) adapter (2)
•
2 N to BNC cables supplied with the Scope Calibrator
Press the NEXT SECTION blue softkey until the display reads “Set up to measure
Pulse Width”. Then follow these steps to calibrate pulse width:
1.
Connect the N to BNC cable supplied with the Scope Calibrator to the Calibrator
Mainframe’s CHAN 1 connector. Connect the other end of the N to BNC cable to
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...