5820A
Service Manual
3-16
one BNC(f) to 3.5 mm(m) adapter then to the DSO’s sampling head through the 3
dB attenuator.
2.
Using the second BNC(f) to 3.5 mm(m) adapter and N to BNC cable, connect the
Calibrator Mainframe’s TRIG OUT connector to the 11801’s Trigger Input.
3.
Set the DSO to these parameters:
•
Main Time Base position (initial):40 ns
•
Vertical scale:
200 mV/div, +900 mV offset
•
Trigger:
source = ext; level = 0.5 V; ext atten = x10; slope = +;
mode = auto
•
Measurement Function:
positive width
4.
Press the GO ON blue softkey.
5.
Adjust the DSO horizontal scale and main time base position until the pulse signal
spans between half and the full display. If no pulse is output, increase the pulse
width using the Calibrator Mainframe front panel knob until a pulse is output.
6.
If prompted to adjust the pulse width by the Calibrator Mainframe display, adjust the
pulse width to as close to the displayed value as possible using the Calibrator
Mainframe front panel knob, then press the GO ON blue softkey.
7.
Allow the DSO width reading to stabilize. Enter the reading via the Calibrator
Mainframe front panel keypad, then press ENTER.
Note
The Calibrator Mainframe issues a warning when the entered value is out
of bounds. If this warning occurs, recheck the setup and carefully re-enter
the reading with the proper multiplier (i.e., m, u, n, p). If the warning still
occurs, enter a value between the displayed pulse width and the previously
entered value. Keep attempting this, moving closer and closer to the
displayed pulse width, until the value is accepted. Complete the pulse width
calibration procedure. The pulse width calibration procedure must now be
repeated until all entered values are accepted the first time without
warning.
8.
Repeat steps 5 to 7 until the Calibrator Mainframe display prompts to connect a
resistor. Press the STORE CONSTS blue softkey to store the new calibration
constants.
3-21. MeasZ Calibration
The MeasZ function is calibrated using resistors and capacitors of known values. The
actual resistance and capacitance values are entered while they are being measure by the
Calibrator Mainframe.
The resistors and capacitor must make a solid connection to a BNC(f) to enable a
connection to the end of the N to BNC cable supplied with the Scope Calibrator. The
resistance and capacitance values must be known at this BNC(f) connector. Fluke uses a
HP 3458A DMM to make a 4-wire ohms measurement at the BNC(f) connector to
determine the actual resistance values and an HP 4192A Impedance Analyzer at 10 MHz
to determine the actual capacitance value.
This procedure uses the following equipment:
•
Resistors of known values: 40
Ω
, 60
Ω
, 600 k
Ω
and 1.5 M
Ω
nominal
•
adapters to connect resistors to BNC(f) connector
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...