5820A
Service Manual
3-34
3-41. Tunnel Diode Pulser Drive Amplitude Verification
This procedure uses the following equipment:
•
Hewlett-Packard 3458A Digital Multimeter
•
BNC(f) to Double Banana adapter
•
N to BNC cable supplied with the Scope Calibrator
Set the Calibrator Mainframe to Edge. Proceed with the following steps:
1.
Connect the Calibrator Mainframe’s CHAN 1 connector to the HP 3458A input,
using the N to BNC cable and the BNC(f) to Double Banana adapter.
2.
Activate the TD Pulser output by pushing the TDPULSE blue softkey. The output
should now be at 80 V peak-to-peak, 100 kHz, STANDBY.
3.
Set the HP 3458A to DCV, NPLC = .001, LEVEL 1, TRIG LEVEL, and the DELAY
to .00012 for measuring the topline and DELAY to .00007 for measuring the
baseline. Manually range lock the HP 3458A to the 100 V dc range.
4.
Change the Calibrator Mainframe output frequency to 10 kHz. Push the operate key,
and use the HP 3458A to measure the topline and baseline.
5.
The peak-to-peak value is the difference between the topline and baseline. Record
these values in Table 3-18, and compare against the listed tolerance.
Table 3-18. Tunnel Diode Pulser Verification
Nominal Value
(V p-p)
Frequency
(Hz)
Measured Value
(V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
11
100
0.2202
11
10000
0.2202
55
100
1.1002
55
10000
1.1002
100
100
2.0002
100
10000
2.0002
3-42. Leveled Sine Wave Amplitude Verification
This procedure uses the following equipment:
•
5790A AC Measurement Standard
•
BNC(f) to Double Banana Plug adapter
•
50
Ω
feedthrough termination
•
N to BNC cable supplied with the Scope Calibrator
Set the Calibrator Mainframe to the Levsine menu. Press
on the Calibrator
Mainframe to activate the output. Then follow these steps to verify the leveled sine wave
amplitude.
1.
Connect the N to BNC cable to the Calibrator Mainframe’s CHAN 1 connector.
Connect the other end of the N to BNC cable to the 50
Ω
feedthrough termination
then to the 5790A INPUT 2 using the BNC(f) to Double Banana adapter.
2.
Set the 5790A to AUTORANGE, digital filter mode to FAST, restart fine, and Hi
Res on.
3.
Program the Calibrator Mainframe to output the voltage listed in Table 3-19.
Содержание 5820A
Страница 12: ...5820A Service Manual viii ...
Страница 14: ...5820A Service Manual 1 2 ...
Страница 28: ...5820A Service Manual 1 16 ...
Страница 30: ...5820A Service Manual 2 2 ...
Страница 38: ...5820A Service Manual 2 10 ...
Страница 102: ...5820A Service Manual 3 64 ...
Страница 104: ...5820A Service Manual 4 2 ...
Страница 110: ...5820A Service Manual 5 2 ...
Страница 140: ...5820A Service Manual 6 2 ...
Страница 144: ...5820A Service Manual 6 6 5820A Final Assembly Sheet 2 of 9 yu173f eps Figure 6 1 Final Assembly cont ...
Страница 146: ...5820A Service Manual 6 8 5820A Final Assembly Sheet 3 of 9 yu174f eps Figure 6 1 Final Assembly cont ...
Страница 148: ...5820A Service Manual 6 10 5820A Final Assembly Sheet 4 of 9 yu175f eps Figure 6 1 Final Assembly cont ...
Страница 152: ...5820A Service Manual 6 14 5820A Final Assembly Sheet 7 of 9 yu177f eps Figure 6 1 Final Assembly cont ...
Страница 154: ...5820A Service Manual 6 16 5820A Final Assembly Sheet 8 of 9 yu178f eps Figure 6 1 Final Assembly cont ...
Страница 156: ...5820A Service Manual 6 18 ...