Installation
Placement and Rack Mounting
2
2-5
•
The air entering the instrument must be room temperature. Make sure that exhaust
from another instrument is not directed into the fan inlet.
•
Clean the air filter every 30 days or more frequently if the 5790A is operated in a
dusty environment. (Instructions for cleaning the air filter are in Chapter 8.)
2-5.
Placement and Rack Mounting
You can place the 5790A on top of a workbench or mount it in a standard 19-inch wide,
24-inch (61-cm) deep equipment rack. For bench-top use, the 5790A is equipped with
non-slipping, non-marring feet. To mount the 5790A in an equipment rack, order the
accessory Y-5737 Rack Mount Kit. The rack mount kit comes with rack ears, 24-inch
slides, fasteners, and an instruction sheet.
2-6.
Accessing the Fuse
Caution
To prevent instrument damage, verify that the correct fuse is
installed in accordance with the line voltage setting.
The line power fuse is accessible on the rear panel. The fuse rating label to the right of
the fuse holder (labeled F1) shows the correct replacement fuse rating for each line
voltage setting. To access the fuse, refer to Figure 2-2 and proceed as follows:
1.
Disconnect line power.
2.
Insert the blade of a standard screwdriver in the slot of the fuse holder labeled F1.
3.
Turn the screwdriver counterclockwise until the cap and fuse pop free.
IEEE-488
CALIBRATION MODE
CALIBRATION STORE
WARNING
GROUNDING CONNECTOR
IN POWER CORD MUST BE
CONNECTED TO ENSURE PROTECTION
FROM ELECTRIC SHOCK.
SERVICE
ENABLE
PERIODIC
NORMAL
VOLTAGE
SELECTION
VOLTAGE
SELECTION
FUSE-F1
FUSE-F1
100V
110V
115V
120V
200V
220V
230V
240V
S2 S3 S4
S2 S3 S4
T 1.5A
250V
(SB)
T .75A
250V
(SB)
RS 232C
CHASSIS
GROUND
S2
S3
S4
FUSE -F1
CAUTION:
FOR FIRE P ROTECTION REPLACE
ONLY WITH A 250V FUSE OF INDICATED RATING.
JOHN FLUKE MFG.CO., INC
MADE IN U.S.A
VARIOUS ASPECTS OF THIS INSTRUMENT ARE PROTECTED BY
ONE OR MORE OF THE FOLLOWING U.S. PATENTS.
4803457, 4860289, 4901052, 4716398, 4847507
NO INTERNAL USER SERVICEABLE
PARTS. REFER SERVICE TO QUALIFIED
SERVICE PERSONNEL
TO CLEAN FILTER, REMOVE
FROM INSTRUMENT AND FLUSH
WITH WARM SOAPY WATER
add3f.eps
Figure 2-2. Accessing the Fuse
Содержание 5790A
Страница 4: ......
Страница 12: ...5790A Operators Manual vi...
Страница 14: ...5790A Operators Manual viii...
Страница 16: ...5790A Operators Manual 1 2...
Страница 34: ...5790A Operators Manual 2 2...
Страница 40: ...5790A Operators Manual 2 8...
Страница 41: ...3 1 Chapter 3 Features Title Page 3 1 Introduction 3 3 3 2 Front Panel Features 3 3 3 3 Rear Panel Features 3 3...
Страница 42: ...5790A Operators Manual 3 2...
Страница 48: ...5790A Operators Manual 3 8...
Страница 108: ...5790A Operators Manual 6 2...
Страница 146: ...5790A Operators Manual 7 2...
Страница 164: ...5790A Operators Manual 7 20...
Страница 166: ...5790A Operators Manual 8 2...
Страница 170: ...5790A Operators Manual 8 6...
Страница 172: ...5790A Operators Manual 9 2...
Страница 176: ...5790A Operators Manual...
Страница 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 186: ...5790A Operators Manual B 2...
Страница 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Страница 188: ...5790A Operators Manual B 4...
Страница 204: ...5790A Operators Manual C 16...
Страница 210: ...5790A Operators Manual D 6...