5790A
Operators Manual
6-4
*SRE
Set Service Request Enable register, described in the Status Information section
below.
*SRE?
Query Service Request Enable register, described in the Status Information section
below.
*STB?
Read status byte. The status byte is described in detail in the instrument status
section below
*TRG
Executes the TRIG command
*TST?
Runs some of the 5790A self tests and returns results. The tests included are: ROM
checksum test, nonvolatile memory test. Also see DIAG.
*WAI
Wait-to-Continue Command. This command prevents further remote commands
from being executed until all previous remote commands have been completely
executed.
Instrument Configuration Commands
CAL_MODE?
Returns the setting of the rear panel CALIBRATION MODE switch.
CAL_SW?
Returns the setting of the rear panel CALIBRATION STORE switch.
CLOCK
Sets the real time clock/calendar
CLOCK?
Queries the value of the real time clock/calendar
DATEFMT
Selects the clock/calendar date format
DATEFMT?
Returns the clock/calendar date format.
EOFSTR
Stores the end of file character sequence for use in cal reports.
EOFSTR?
Returns the end of file character sequence set for cal reports.
EXTGUARD
Sets external guard.
EXTGUARD?
Returns the setting of external guard.
EXTRIG
Selects single trigger mode.
EXTRIG?
Returns the setting of EXTRIG.
FIRSTIN
Stores the power-up configuration for the initial input.
FIRSTIN?
Returns the power-up configuration for the initial input.
FORMAT
Formats the EEPROM device. This command stores default values into the
EPROM, overwriting all existing values. Use with extreme care.
SP_SET
Programs serial port nonvolatile settings.
SP_SET?
Returns the serial port nonvolatile settings.
Measurement and Transfer Commands
DELTA?
Returns the delta (transfer) value from the 5790A.
DFILT
Sets the digital filter parameters for mode and restart.
DFILT?
Returns the digital filter parameters for mode and restart.
DUNIT
Sets the delta unit value.
DUNIT?
Returns the delta unit value.
HIRES
Enables and disables higher resolution amplitude display.
Содержание 5790A
Страница 4: ......
Страница 12: ...5790A Operators Manual vi...
Страница 14: ...5790A Operators Manual viii...
Страница 16: ...5790A Operators Manual 1 2...
Страница 34: ...5790A Operators Manual 2 2...
Страница 40: ...5790A Operators Manual 2 8...
Страница 41: ...3 1 Chapter 3 Features Title Page 3 1 Introduction 3 3 3 2 Front Panel Features 3 3 3 3 Rear Panel Features 3 3...
Страница 42: ...5790A Operators Manual 3 2...
Страница 48: ...5790A Operators Manual 3 8...
Страница 108: ...5790A Operators Manual 6 2...
Страница 146: ...5790A Operators Manual 7 2...
Страница 164: ...5790A Operators Manual 7 20...
Страница 166: ...5790A Operators Manual 8 2...
Страница 170: ...5790A Operators Manual 8 6...
Страница 172: ...5790A Operators Manual 9 2...
Страница 176: ...5790A Operators Manual...
Страница 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 186: ...5790A Operators Manual B 2...
Страница 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Страница 188: ...5790A Operators Manual B 4...
Страница 204: ...5790A Operators Manual C 16...
Страница 210: ...5790A Operators Manual D 6...