Remote Commands
Alphabetical List of IEEE-488.2 Common Commands
6
6-7
CAL_NEXT
Initiates the next calibration point.
CAL_NEXT?
Returns the value of the next point to be calibrated when in the middle of calibrating
a range.
CAL_OFF
Cancels any interactive calibration underway (that is, one started with the CAL_AC,
CAL_I2, and CAL_DC commands).
CAL_SHIFT?
Returns the calibration shift for a single calibration point.
CAL_SKIP
Advances to the next step of an interactive procedure without executing the step
(like a CAL_NEXT except the step isn't executed).
CAL_SLST?
Returns the calibration shifts for a group of service calibration points.
CAL_SPEC?
Returns the uncertainty specification for a specified input value, range, and
frequency.
CAL_STLST?
Lists the calibration constant groups that have been updated but not saved (and
therefore, how many would be written to nonvolatile memory by a CAL_STORE
command with the same argument).
CAL_STORE
Stores calibration constants in the specified group list into nonvolatile memory.
CAL_STORE?
Tells how many calibration constant groups have been modified but not saved (this
is like CAL_STLST? except only the initial number is given).
CAL_TEMP
Sets the temperature for calibration. This should be done before CAL_DC, CAL_I2,
and CAL_AC commands. Once set, the temperature is used for all calibration
activities until changed. If not set before a calibration activity is complete, the 5790A
uses a default of 23.0
°
C.
CAL_TEMP?
Returns the most recent calibration temperature for a specified type of calibration.
CAL_ZERO
Executes internal dc zeros calibration.
CMDSTR?
Returns the command string that caused the last error.
DIAG
Run a self diagnostic routine.
DIAGFLT
Set the 5790A response to errors that occur during execution of diagnostics under
remote control. (Nonvolatile.)
DIAGFLT?
Returns the setting of the response to errors that occur during diagnostics executed
under remote control.
ERR?
Returns the first error code and description from the 5790A error queue. This error
is then taken off the queue.
EXPLAIN?
Explains an error code. (5700A compatibility.)
FAULT?
Returns an error code and takes the error off the error queue. (5700A compatibility.)
6-2.
Alphabetical List of IEEE-488.2 Common Commands
*CLS
(Clear status.) Clears the ESR, ISCR0 and ISCR1, the error queue, and the RQS bit in the
status byte. This command terminates a pending operation complete (*OPC).
Parameter
None
Содержание 5790A
Страница 4: ......
Страница 12: ...5790A Operators Manual vi...
Страница 14: ...5790A Operators Manual viii...
Страница 16: ...5790A Operators Manual 1 2...
Страница 34: ...5790A Operators Manual 2 2...
Страница 40: ...5790A Operators Manual 2 8...
Страница 41: ...3 1 Chapter 3 Features Title Page 3 1 Introduction 3 3 3 2 Front Panel Features 3 3 3 3 Rear Panel Features 3 3...
Страница 42: ...5790A Operators Manual 3 2...
Страница 48: ...5790A Operators Manual 3 8...
Страница 108: ...5790A Operators Manual 6 2...
Страница 146: ...5790A Operators Manual 7 2...
Страница 164: ...5790A Operators Manual 7 20...
Страница 166: ...5790A Operators Manual 8 2...
Страница 170: ...5790A Operators Manual 8 6...
Страница 172: ...5790A Operators Manual 9 2...
Страница 176: ...5790A Operators Manual...
Страница 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 186: ...5790A Operators Manual B 2...
Страница 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Страница 188: ...5790A Operators Manual B 4...
Страница 204: ...5790A Operators Manual C 16...
Страница 210: ...5790A Operators Manual D 6...