5790A
Operators Manual
6-14
Example
If you send the compound command "ABORT;*OPC?", the *OPC? is not waiting for
ABORT to complete, it is waiting for the next measurement to complete.
CAL_AC
Begins an interactive ac calibration procedure for one range or all ranges. This does not
do a calibration step, but sets up the 5790A so the next CAL_NEXT command does the
first step and the next CAL_NEXT? command returns the first step.
Parameters
1.
INPUT1 or INPUT2 (The input to be used, signifies normal ac calibration), or
WBND for Option -03 Wideband flatness calibration.
2.
(Optional) A number that defines the range amplitude (same as for the RANGE
command). If this parameter is included, do only the specified range; if not
included, do all ranges.
3.
(Optional only if parameter 2 is present; prohibited otherwise) number that
specifies which step to jump to of the procedure for the specified range.
Example
The sequence of commands for calibrating an entire range is:
CAL_TEMP 23.5 (optional)
CAL_AC <input>, <amplitude> input is location, amplitude tells range
CAL_NEXT <amplitude> this is the first step (frequency is implicit)
CAL_NEXT <amplitude> this is the last step
An error occurs if either the entered amplitude is out of range or the measured input is out
of range. The CAL_NEXT? command gives the nominal amplitude and frequency for the
next CAL_NEXT command and returns 0,-1 after the last CAL_NEXT command has
been done for a range.
CAL_AC?
Lists the steps of an interactive calibration procedure or a portion of a procedure as it
would run when requested with a CAL_AC command with the same parameters.
Parameters
1.
INPUT1 or INPUT2 (The input to be used, signifies ac calibration), or WBND
for Option -03 Wideband calibration
2.
(Optional) A number that defines the range amplitude (same as for the RANGE
command). If this parameter is included, list only the steps for the specified
range; if not included, list steps for all ranges.
3.
(Optional only if parameter 2 is present; prohibited otherwise) A number that
specifies which step to jump to to begin listing steps for the specified range.
Response
(IAD) A list in the following form:
<# of ranges><EOL>
<range 1 max>,<# of steps in range><EOL>
<1st step amplitude in volts>,<1st step frequency in Hz><EOL>
...
Содержание 5790A
Страница 4: ......
Страница 12: ...5790A Operators Manual vi...
Страница 14: ...5790A Operators Manual viii...
Страница 16: ...5790A Operators Manual 1 2...
Страница 34: ...5790A Operators Manual 2 2...
Страница 40: ...5790A Operators Manual 2 8...
Страница 41: ...3 1 Chapter 3 Features Title Page 3 1 Introduction 3 3 3 2 Front Panel Features 3 3 3 3 Rear Panel Features 3 3...
Страница 42: ...5790A Operators Manual 3 2...
Страница 48: ...5790A Operators Manual 3 8...
Страница 108: ...5790A Operators Manual 6 2...
Страница 146: ...5790A Operators Manual 7 2...
Страница 164: ...5790A Operators Manual 7 20...
Страница 166: ...5790A Operators Manual 8 2...
Страница 170: ...5790A Operators Manual 8 6...
Страница 172: ...5790A Operators Manual 9 2...
Страница 176: ...5790A Operators Manual...
Страница 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 186: ...5790A Operators Manual B 2...
Страница 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Страница 188: ...5790A Operators Manual B 4...
Страница 204: ...5790A Operators Manual C 16...
Страница 210: ...5790A Operators Manual D 6...