5790A
Operators Manual
6-6
ISCE0
Sets the Instrument Status One To Zero Change Enable register
ISCE0?
Returns the contents of the Instrument Status Zero To One Change Enable register.
ISCE1
Sets the Instrument Status Zero To One Change Enable register.
ISCE1?
Returns the contents of the Instrument Status One To Zero Change Enable register.
ISCR0?
Returns the contents of the Instrument Status One To Zero Change Register.
ISCR1?
Returns the contents of the Instrument Status Zero To One Change Register.
ISR?
Returns the contents of the Instrument Status Register.
MODE?
Returns the operating mode (MEASUREMENT, CALIBRATION, DIAGNOSTIC,
CALWAITING).
MODESTR?
Returns a string that describes what a calibration or diagnostic procedure is doing.
ONTIME?
Returns the time in minutes since the power was turned on.
5790A Calibration, Testing, and Diagnostics Commands
ABORT
Aborts a measurement, calibration, or diagnostic procedure, or report being printed
CAL_AC
Begins an interactive ac calibration procedure for an ac range.
CAL_AC?
Lists the steps of an interactive calibration procedure or a portion of a procedure as
it would run when requested with a CAL_AC command with the same parameters.
CAL_BACKUP
Makes it so that the next CAL_NEXT command redoes the previously finished
calibration step.
CAL_CLST?
Lists the symbolic names for all the calibration constants for a specified range.
CAL_CONST?
Returns the value of a particular calibration constant.
CAL_DATE?
Returns the date of the most recent calibration.
CAL_DAYS?
Returns the number of days elapsed in the calibration interval.
CAL_DC
Begins a calibration procedure for a dc range. As in CAL_AC, this merely specifies
which step the next CAL_NEXT command does.
CAL_DC?
Lists the steps of an interactive calibration procedure or a portion of a procedure as
it would run when requested with a CAL_DC command with the same parameters.
CAL_FPT
Corrects the reference given for an ac calibration step after the fact, and as a result
change the resulting calibration constant or constants.
CAL_FPT?
Supplies the first two parameters to be used with a CAL_FPT to correct the next
calibration step to be performed.
CAL_FREQ
Executes the frequency calibration procedure. An error is generated if the frequency
or amplitude measurement is out of range. If an error occurs, the frequency
calibration procedure is not executed.
CAL_I2.
Begins the INPUT 2 vs. INPUT 1 calibration procedure.
CAL_I2?
Lists the steps in the INPUT 2 vs. INPUT 1 calibration procedure.
CAL_INPUT?
Returns the input to which the reference will be applied for the next calibration step.
CAL_INTV
Sets the calibration interval. This is stored in nonvolatile memory and used to
calculate the output uncertainty.
CAL_INTV?
Returns the calibration interval.
Содержание 5790A
Страница 4: ......
Страница 12: ...5790A Operators Manual vi...
Страница 14: ...5790A Operators Manual viii...
Страница 16: ...5790A Operators Manual 1 2...
Страница 34: ...5790A Operators Manual 2 2...
Страница 40: ...5790A Operators Manual 2 8...
Страница 41: ...3 1 Chapter 3 Features Title Page 3 1 Introduction 3 3 3 2 Front Panel Features 3 3 3 3 Rear Panel Features 3 3...
Страница 42: ...5790A Operators Manual 3 2...
Страница 48: ...5790A Operators Manual 3 8...
Страница 108: ...5790A Operators Manual 6 2...
Страница 146: ...5790A Operators Manual 7 2...
Страница 164: ...5790A Operators Manual 7 20...
Страница 166: ...5790A Operators Manual 8 2...
Страница 170: ...5790A Operators Manual 8 6...
Страница 172: ...5790A Operators Manual 9 2...
Страница 176: ...5790A Operators Manual...
Страница 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 186: ...5790A Operators Manual B 2...
Страница 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Страница 188: ...5790A Operators Manual B 4...
Страница 204: ...5790A Operators Manual C 16...
Страница 210: ...5790A Operators Manual D 6...