5790A
Operators Manual
1-6
1-10.
Current Shunt Adapters
Accessory 5790A-7001 allows you to connect Fluke A40 or A40A Current Shunts the
5790A SHUNT input for making current transfers up to 20 A. Accessory 792A-7004
allows you to connect Fluke A40 Current Shunts to the INPUT 1 Type "N" connector for
making current transfers up to 5 A.
1-11.
Warranty and Service Information
Each 5790A is warranted for a period of 1 year on delivery to the original purchaser. The
warranty is printed on the reverse side of the title page of this manual.
Factory authorized service for the 5790A is available at selected Technical Service
Centers. For warranty or after-warranty service, contact the nearest Technical Service
Center for instructions. A complete list of Technical Service Centers is provided in
Appendix D.
To reship the 5790A, use its original shipping carton. If the original carton is not
available, use a container that provides adequate protection during shipment. Protect the
5790A with at least three inches of shock-absorbing material on all sides of the container.
Do not use loose fill to pad the shipping container. Loose fill allows the instrument to
settle to one corner of the shipping container, which could result in damage during
shipment.
1-12.
Specifications
Specifications are valid after allowing a warm-up period of 30 minutes, or twice the time
the 5790A has been turned off, whichever is less. For example, if the 5790A has been
turned off for 5 minutes, the warm-up period is 10 minutes. To simplify evaluation of
how the 5790A covers you workload, use the Absolute Uncertainty Specification. Those
include stability, temperature coefficient, linearity, and traceability to external standards.
Note
When you use the 5790A within ±5 °C of the temperature of the last
calibration, you do not need to add anything to the Absolute Uncertainty
Specifications to determine the ratios between 5790A uncertainties and the
uncertainties of a unit under test. The initial calibration at Fluke is done at
23 °C. (You can verify the temperature of the last calibration at any time by
pressing the [SPEC] key.)
Use the Relative Uncertainty Specifications if you use a different procedure to calibrate
the 5790A than is specified in the 5790A Service Manual. To calculate absolute
uncertainty specifications under such conditions, combine the absolute uncertainty
associated with your external equipment and calibration procedures with the Relative
Uncertainty Specifications. Secondary Performance and Operating Characteristics are
provided for special calibration requirements such as stability or operation at temperature
extremes.
Содержание 5790A
Страница 4: ......
Страница 12: ...5790A Operators Manual vi...
Страница 14: ...5790A Operators Manual viii...
Страница 16: ...5790A Operators Manual 1 2...
Страница 34: ...5790A Operators Manual 2 2...
Страница 40: ...5790A Operators Manual 2 8...
Страница 41: ...3 1 Chapter 3 Features Title Page 3 1 Introduction 3 3 3 2 Front Panel Features 3 3 3 3 Rear Panel Features 3 3...
Страница 42: ...5790A Operators Manual 3 2...
Страница 48: ...5790A Operators Manual 3 8...
Страница 108: ...5790A Operators Manual 6 2...
Страница 146: ...5790A Operators Manual 7 2...
Страница 164: ...5790A Operators Manual 7 20...
Страница 166: ...5790A Operators Manual 8 2...
Страница 170: ...5790A Operators Manual 8 6...
Страница 172: ...5790A Operators Manual 9 2...
Страница 176: ...5790A Operators Manual...
Страница 185: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Страница 186: ...5790A Operators Manual B 2...
Страница 187: ...Appendices ASCII and IEEE 488 Bus Codes B B 3...
Страница 188: ...5790A Operators Manual B 4...
Страница 204: ...5790A Operators Manual C 16...
Страница 210: ...5790A Operators Manual D 6...