medTester 5000C
Operators Manual
5-10
ECG Performance Waveforms
If you have the Waveform/Extended Testing module installed on your
medTester 5000C, and you have performed a lead test, this message appears:
This is a prompt to output an autosequence of performance waves to a
patient monitor from the patient leads attached to the medTester 5000C.
Waveforms are output in the order described in the table below for various
periods of time.
Waveform Name
Test Duration
(In Seconds)
2 Hz Square
Continuous
4-second Pulse
8
10 Hz Sine
4
40 Hz Sine
4
60 Hz Sine
60
100 Hz Sine
4
1 kHz Square
4
2 Hz Triangle
Continuous
CMRR 60 Hz
Continuous
ECG 60 BPM
Continuous
ECG 30 BPM
30
ECG 60 BPM
30
ECG 120 BPM
30
ECG 240 BPM
30
Note
Tests with a predefined duration automatically end. However,
you can choose to terminate them at any time by pressing
NEXT
.
Those marked
continuous
can be terminated by pressing
NEXT
.
OUTPUT PERFORMANCE WAVES?
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...