Biomedical Test System
SPO2 Module
15
15-25
Note
If you choose "Not Specified", you can keep it generic.
However, you must manually set the SPO2 simulator to the
manufacturer being tested at run time. Press
F5 OK
to enter your
selection.
3.
SETTLING TIME: Enter the number of heartbeats to wait after
starting a simulation, for the pulse oximeter to settle its readings
(0-99).
4.
HEART RATE +/- LIMIT: Enter the limit of variation allowed
in Beats Per Minute for heart rates for the SPO2 simulation steps.
00 BPM turns off limit checking.
5.
The following parameters can be set for up to 10 SPO2
simulation steps for all analyzers except the Index 2XL, for which
15 steps can be programmed:
•
SPO2 LEVEL: Enter the percentage of oxygen saturation desired:
For the Oxitest Plus, select from discrete levels of 55%, 70%, 80%,
90%, 93%, and 97%; for the CardioSat 100 and Index 2XL, enter
any level from 35% to 100%.
•
SPO2 +/- LIMIT: Enter the limit of variation allowed in percent
for SPO2 for this SPO2 simulation step. 00% turns off limit
checking.
•
HEART RATE: Enter the heart rate, for the Oxitest Plus, from 20
to 250 BPM; for the CardioSat 100 and Index 2XL, from 30 to 250
BPM.
•
AMPLITUDE: Enter the amplitude in percentage of full amplitude,
from 0% to 100%.
•
TLC (Index 2XL only): Enter the Transmission Level Control value
from 0000 to 4095. The TLC value defaults to the “Base TLC” level
for the make of SPO2 manufacturer found in the previous table.
•
ARTIFACT (Oxitest Plus and Index 2XL only):
-
For Oxitest the selections are: Off, Movement, Spike or
Tremor.
-
For the Index 2XL the selections are: Off, Level 1 or Level 2.
Each of these levels contain eight different motion tests. If you
choose Level 1 or 2, then you will be prompted to enter the
MOTION TEST SPO2 +/- LIMIT
(defaults to 04%).
After each step selection, you can press
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...