Biomedical Test System
medCheck Module
11
11-5
As you can see, some checkitems appear in the medTester as prompts to the
operator to do something, for example, a physical inspection of the EUT or a
check of some function. Other checkitems are autosequences. Autosequence
46 is a customized (user defined) autosequence that can execute a variety of
performance tests on this defibrillator. For more information about
defibrillator autosequences, see Chapter 9,
Defibrillator Module
. To learn more
about customizing autosequences, see Chapter 6,
Customize Your
medTester 5000C
.
This checklist could be downloaded to the medTester with a list of control
numbers. Control numbers identify unique pieces of equipment. Each control
number has a checklist assigned to it. When you perform the checklist on a
control number-referenced piece of equipment, the test record that is
generated includes the control number. If you test a defibrillator of this model
type and it has no control number, the medTester can prompt you to assign
one. This lets you bring new equipment into inventory as you test. Of course,
you still have the option of making an equipment record and assigning a
control number prior to testing it with the checklist.
In this example the last checkitem in the checklist prompts the operator to
write out a preventive maintenance tag that is placed on the EUT after the
checklist runs. The medTester 5000C also has an automated tag generation
capability. This requires the connection of an optional test label printer. See
the
Test Tag Printer
section in Chapter 5.
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...