medTester 5000C
Operators Manual
1-12
medTester 5000C Instrument Specifications
Line Voltage and Measurements
•
Hot to Neutral.
•
Neutral to Ground.
•
Hot to Ground.
•
Range: 200.0 V RMS.
•
Accuracy: ±5% of range.
Leakage Current
Leakage current is measured through a 1 k
Ω
AAMI load RMS or DC
measured and displayed in microamperes. These tests can be thought of as
measuring the voltage, in millivolts across the same load.
Leakage Current Measurements
•
Case external.
•
Case internal.
•
External.
•
ECG—Leads are RL, RA, LA, LL, and V1-V6 tied together:
1.
To ground, all leads or individual leads.
2.
Lead to lead, individual leads to all other leads together.
3.
Lead isolation (RMS only), all leads or individual leads.
This test is with line voltage applied from the lead(s) to
ground, current limited by 120 k
Ω
.
Note
If you receive abnormally high leakage current measurements
with V1-V6 leads tied together, you can inspect individual V
leads by removing the other leads. This procedure allows you to
isolate the lead with a high leakage current reading.
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...