Biomedical Test System
The Wedge Adapter
13
13-9
Configuring an Autosequence Port
You can execute defibrillator, IV pump, ESU, and SPO2 autosequences
through a Wedge port in the same manner that you can execute them through
the medTester 5000C’s COM2 port. This allows you, for example, to connect
both a defibrillator analyzer such as the Impulse 3000 or the Impulse 4000
and an IV pump tester such as the IDA 4 to defined ports on the Wedge.
Then whenever you execute one of these autosequences, the
medTester 5000C knows to which port it must send the autosequence.
The medTester 5000C also uses the configured autosequence port for
executing any autosequence that is embedded as a checkitem in a checklist.
For example, if the medTester executes a checklist that contains a defibrillator
autosequence, it directs that autosequence to the defibrillator autosequence
port.
Each autosequence type has a default port setting. Defibrillator autosequences
are assigned to the Wedge’s COM2-2 port. Follow these instructions to
configure a different port for defibrillator autosequences:
1.
From the medTester 5000C main menu,
MENU 1
, press:
UTIL
F5
2.
From the utility menu, press the medTester right arrow and enter
the command:
WPORTS
F5
3.
Press:
DEFIB
F2
to see this display:
4.
Press the medTester up arrow key to increment the number to the
port you want as the defibrillator autosequence port, or press the
down arrow key to decrement the number.
5.
After selecting the normal port, press
F5
to store your selection.
SET WEDGE PORT FOR DEFIB: 2
STORE
F5
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...