medTester 5000C
Operators Manual
16-8
Demand Test
The demand test checks that in the demand mode, the pacer will pulse
interactively with ECG waveforms at different rates. The demand test
requires that the pacer is in the demand mode and that the pacer ECG
monitor leads are connected to the ECG posts of the Impulse 4000.
Underdrive
The demand underdrive test checks to make sure the pacer outputs pacing
pulses when the ECG signal is too slow to sustain a patient without pacing.
The medTester displays the pacing rate and the underdrive ECG rate. Set the
pacer to demand mode at the pacing rate and turn on pacing. Press F5
START.
The medTester turns on the ECG waveform on the Impulse 4000.
The medTester indicates that pacing should be on. Observe the pacer ECG
monitor to check that the ECG signal is running and that pacing is on.
Press F4 PASS or F5 FAIL to store that with the record. FAIL is stored with
a “#”.
Overdrive
The demand overdrive test checks to make sure the pacer does not output
pacing pulses when the ECG signal is fast enough that the patient does not
need to be paced.
The medTester displays the pacing rate and the overdrive ECG rate. Set the
pacer to demand mode at the pacing rate and turn on pacing. Press F5
START.
The medTester turns on the ECG waveform on the Impulse 4000.
The medTester indicates that pacing should be off. Observe the pacer ECG
monitor to check that the ECG signal is running and that pacing is off.
Press F4 PASS or F5 FAIL to store that with the record. FAIL is stored with
a “#”.
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...