medTester 5000C
Operators Manual
12-14
MEASUREMENT COMMANDS
Note: These commands set up a measurement, according to global variables as needed, and
after a 0.5 second settling time, return the measurement reading. The measurement remains set
up, so that additional readings can be taken with the REPEAT command. Global variables can
be changed while the measurement remains set up, then more readings can again be taken
with the REPEAT command. A measurement can be turned off by any non-measurement/non-
setup command or by the EXIT command. While a measurement is set up, a new measurement
command is allowed without previously exiting. The units of measurement are listed in
parentheses. For ECG lead isolation tests, the isolation voltage stays on only during each
measurement, then turns off. A REPEAT command would turn it back on as long as needed for
the next measurement.
Command
Explanation
The response is indicated with the unit of measurement listed in
parentheses.
SV12
System voltage, hot to neutral.
(V RMS)
SV1G
System voltage, neutral to ground.
(V RMS)
SV2G
System voltage, hot to ground.
(V RMS)
CLKE
Case leakage current, external leads.
(µA RMS or DC)
CLKI
Case leakage current, internal leads.
(µA RMS or DC)
XTMI
External meter leakage current.
(µA)
ALGD
ECG leakage current, All leads to ground.
(µA RMS or DC)
RLGD
ECG leakage current, Right Leg to ground.
(µA RMS or DC)
RAGD
ECG leakage current, Right Arm to ground.
(µA RMS or DC)
LAGD
ECG leakage current, Left Arm to ground.
(µA RMS or DC)
LLGD
ECG leakage current, Left Leg to ground.
(µA RMS or DC)
VGND
ECG leakage current, V leads to ground.
(µA RMS or DC)
RLLD
ECG leakage current, Right Leg to leads.
(µA RMS or DC)
RALD
ECG leakage current, Right Arm to leads.
Содержание medTester 5000C Operators
Страница 6: ......
Страница 32: ...medTester 5000C Operators Manual 1 20...
Страница 40: ...medTester 5000C Operators Manual 2 8...
Страница 46: ...medTester 5000C Operators Manual 3 6...
Страница 59: ...Biomedical Test System Manual Tests 4 4 13 Figure 4 1 Three Types of Line Voltage Measurements...
Страница 60: ...medTester 5000C Operators Manual 4 14 Figure 4 2 Case External Leakage...
Страница 61: ...Biomedical Test System Manual Tests 4 4 15 Figure 4 3 Case Internal Leakage...
Страница 62: ...medTester 5000C Operators Manual 4 16 Figure 4 4 External Leakage Current...
Страница 64: ...medTester 5000C Operators Manual 4 18 Figure 4 6 ECG Interlead Leakage...
Страница 65: ...Biomedical Test System Manual Tests 4 4 19 Figure 4 7 ECG Lead Isolation Test All Leads and Individual Leads...
Страница 66: ...medTester 5000C Operators Manual 4 20 Figure 4 8 Equipment Load Current...
Страница 67: ...Biomedical Test System Manual Tests 4 4 21 Figure 4 9 Power Cord Resistance Test...
Страница 68: ...medTester 5000C Operators Manual 4 22 Figure 4 10 External Resistance...
Страница 69: ...Biomedical Test System Manual Tests 4 4 23 Figure 4 11 Isolated Power System and Ground Fault Test...
Страница 70: ...medTester 5000C Operators Manual 4 24...
Страница 108: ...medTester 5000C Operators Manual 6 16...
Страница 114: ...medTester 5000C Operators Manual 7 6...
Страница 124: ...medTester 5000C Operators Manual 8 10...
Страница 178: ...medTester 5000C Operators Manual 11 14...
Страница 196: ...medTester 5000C Operators Manual 12 18...
Страница 210: ...medTester 5000C Operators Manual 13 14...
Страница 256: ...medTester 5000C Operators Manual 15 28...
Страница 274: ...medTester 5000C Operators Manual 16 18...
Страница 291: ...Biomedical Test System Index Index 3 installing 13 4 operating 13 7 setting baud rates 13 11 setting output ports 13 13...
Страница 292: ...medTester 5000C Operators Manual Index 4...